Moxa NPort 5110A and Ohaus Defender 3000 scale

How to connect your scale to a network (Ethernet) using Moxa’s NPort 5110A

Many scales and balances are equipped with an RS-232 interface. Moxa’s NPort series and similar serial device servers allow you to connect such scales to a network (Ethernet). In this article, I’ll show you how to connect a Moxa NPort 5110A to an Ohaus scale and how to configure and test it.

1. Physical connection

The Moxa Nport 5110A has an RS-232 port (DE9M) on one side and an Ethernet port on the other side:

Data cables (red) are not included.

To communicate with your scale, you’ll need a matching RS-232 cable. The Ohaus Defender 3000 scale used in this example requires a DE9M to DE9F straight (1:1) cable:

Ohaus scale with Ethernet connection through Moxa NPort 5110A

Once the physical connection to the scale has been established, the Moxa NPort has to be configured. There are several ways to do so. I prefer the following method:

  1. Connect the NPort directly to a single computer (not a LAN) with an Ethernet cable.
  2. Use the NPort Administator software included with the device to configure it.

2. Network configuration

Start the NPort Administrator and locate your NPort by clicking on the Search button:NPort Administrator: Search

It should be found at the default IP address (192.168.127.254).

Note: It’s not always necessary to change your PC’s IP address if you’re running the NPort Administrator software. However, if you receive error messages during the following steps, try to temporarily set your computer’s IP address to an address on the same subnet (such as 192.168.127.1).

Select the Ethernet connection under Control Panel > Network and Internet > Network connections, right-click on it and choose Properties, then select Internet Protocol Version 4 (TCP/IPv4) and click on Properties. Make a note of your current settings so that you can restore them later, then enter the temporary IP address:
Ethernet IP address setting in Windows 10

If the NPort is shown as locked, right-click on it and select Unlock:
NPort Administrator: UnlockThe default password is “moxa”.

Right-click again and select Configure, then select the Network tab:
NPort Administrator: Network configurationEnter a static IP address, subnet mask and gateway that work on your network (or select DHCP or BOOTP for automatic address assignment, though this is not recommended for most operating modes).

To make the NPort accessible on our network, I set its IP address to 192.168.0.253:
NPort Administrator: Modified network settings

With the network configuration complete, you can disconnect the NPort 5110A from your computer and connect it to your network (LAN).

3. RS-232 configuration

It is essential that both the NPort and the scale use the same settings for the serial port (RS-232). In the Serial tab, select the port (the NPort 5110A only has one, but you still have to select it), check Modify and click on Settings:
NPort Administrator: Default serial settings

Enter the RS-232 settings found in your scale’s menu or user manual. The default settings of our Ohaus Defender 3000 scale are shown in the screenshot below:NPort Administrator: Serial configuration (RS-232)

Important: By default, the NPort is set to CTS/RTS flow control! You cannot change this setting when using the Setup Wizard through the web console. As mentioned before, I recommend using the NPort Administrator software instead.

4. Set an operating mode

To change the mode, go to the Operating Mode tab, check Modify, select the port and click on Settings:NPort Administrator: Operating Mode

The NPort 5110A supports several operating modes, I will only discuss the following two here:

  • Real COM mode: This mode allows you to create a virtual COM port on a PC and use the scale as if it were connected directly to that PC. This is very useful when you’re using software which only supports connections to COM ports and cannot communicate over TCP/IP (such as our 232key virtual keyboard wedge software).
    Please note that you’ll have to install a virtual COM port driver on the PC that communicates with the scale. In my tests on Windows 10, setting up the COM port through NPort Administrator did not create a new virtual COM port on the system. However, using the NPort Windows Driver Manager worked.
  • TCP Server: The NPort acts as a server, waiting for incoming connections from TCP clients on port 4001 (or on another user-defined port). The maximum number of concurrent connections can be changed from 1 (default) to up to 8. This mode works with software like our Simple Data Logger and many others.
    NPort Administartor: TCP Server mode

    Note: It is not necessary to set the data packing options unless you want to optimize either for minimal latency or maximum throughput. By default, the device will try to find a reasonable compromise (this is likely achieved by observing the delay between the data received over RS-232 to determine when a “line” of data is complete).

    When I set the scale to continuous transmission mode (wich has no delay between the weight values), the NPort packed 104 bytes in each Ethernet frame, corresponding to almost 6 weight values (the scale sends 80 values/s):
    Wireshark screenshot

For further details and for information on the other modes, please consult the user manual.

5. Run a test

Depending on the operating mode chosen above, you’ll have to use different programs to test the NPort 5110A. In addition to the software I’ve already mentioned above, you could use the following:

Real COM mode

A terminal program like Termite or HTerm can connect to the virtual COM port. Make sure that the connection parameters correspond to the settings of your scale.

TCP Server

You can use PuTTY set to “Raw” TCP mode to act as a TCP client:
PuTTY configurationThe following screenshot shows commands sent to the scale and the replies:PuTTY communication with Ohaus scaleAnother option is our free TCPTester software. It repeatedly sends a user-defined command to the scale (e.g. to request the weight). Once a reply has been received, the command is repeated:

TCPTester tcp utility used to test Moxa NPort connected to Ohaus scale
TCPTester free TCP test utility available from Smartlux

During testing, it can be useful to observe the LEDs on the NPort:

  • The “Ready” LED should be green.
  • The “Link” LED is green when the NPort has established a 100 Mbs Ethernet connection and orange on a 10 Mbps connection.
  • The Tx/Rx LED flashes orange when the NPort receives data from the scale and flashes green when the NPort is sends data to the scale over the serial port. Therefore, If you’re sending commands to a scale, you should be able to see it flash green. A reply from the scale would cause the LED to flash orange. If this is too difficult to see, consider using the web console to diagnose connection issues (I found it worked best in Firefox):
    Moxa web console monitor async

I hope this article has been useful. What are your experiences with Moxa’s NPort serial device servers? Please don’t hesitate to leave a comment! For support, please contact Moxa or your vendor.

Did you find this article helpful? You can support this blog by purchasing a PDF version for US$2/copy (contains no additional information).

Leave a Reply to Mark Letheren Cancel reply

Your email address will not be published. Required fields are marked *

The maximum upload file size: 12 MB. You can upload: image, text. Drop file here

This site uses Akismet to reduce spam. Learn how your comment data is processed.

4 thoughts on “How to connect your scale to a network (Ethernet) using Moxa’s NPort 5110A”

  1. HI there, I stumbled across your blog post. We have developed an internal shipping system, but none the scales can be communicated with, so we are always “keying” in the weight.
    Our shipping system runs in a web browser, but is developed in Visual Studio C# ASP.NET MVC .. would your product work with this platform?

    1. Hi, if your backend is deployed locally and if you can create a TCP connection as a client, you could use the setup described in this article (with the Moxa NPort 5110A acting as a TCP server).

      A different approach would be to use our 232key software to enter the weight as keystrokes into you existing browser-based application (which would not have to be modified).