Solution: Err 8.2 on Ohaus Defender scales

If your new Ohaus Defender shows “Err 8.2”, you have probably not yet removed the shipping spacers. These protect your scale from damage during transport, but have to be taken out before use.

1. Remove the stainless stainless steel platform pan. You’ll see 4 red plastic shipping spacers wedged between the overload protection screws and the upper frame of the platform:
Ohaus Defender 3000 shipping spacer

2. Remove these spacers by hand or with an adequate tool (without turning the overload protection screws):
Ohaus Defender shipping spacers removal

3. Place the stainless steel pan on the platform. Your scale is ready to weigh.

Leave a Reply

Your email address will not be published. Required fields are marked *

The maximum upload file size: 12 MB. You can upload: image, text. Drop file here

This site uses Akismet to reduce spam. Learn how your comment data is processed.

7 thoughts on “Solution: Err 8.2 on Ohaus Defender scales”

  1. sir I urgently need you advice to clear the clear on 2000series ohaus weighing digital balance. I have follow Stephen advice yet it didn’t clear I have being using it for two years ago but flood overflow submerged the base thereafter this error continually displayed. what can I do. thanks

  2. My spacers were removed however my scale still shows this error after a year of usage. IS there a way to find out what is causing it?

    1. Err 8.2 means “Weight reading below «Power On Zero» limit.” The service manual says you should add a load to the scale or reset this limit by calibrating the scale (if you have the necessary calibration weights).